Description

Compact device with a set of removable dynamic and ultrasonic contact impedance probes is designed to measure the hardness and temporary resistance (σ) of different metals: low-carbon steel, cast iron, stainless and high alloyed steels, non-ferrous metals.

The device combines two different physical methods in one uniform operation concept - Quasi-Static Ultrasonic Contact Impedance (UCI) and the Dynamic Rebound Hardness Testing methods.





Technical features

  1. Measurement range:
    • according to Brinell hardness scale, HB, 75...650;
    • according to Rockwell hardness scale, HRC 20...70;
    • according to Vickers hardness scale, HV 75...1000;
  2. Temporal resistance, MPa: 370… 1740;
  3. Basic measurement error:
    • HB, not more than:
      • in range of (75… 150)    +/-10;
      • in range of (150… 300)    +/-15;
      • in range of (300… 650)    +/-20;
    • HRC, not more than +/-2;
    • HV, not more than:
      • in range of (240… 500)    +/-15;
      • in range of (500… 800)    +/-20;
      • in range of (800… 940)    +/-25;
  4. Number of stored individual calibrations for each scale of hardness - up to 5;
  5. Number of memory cells for measurement results - up to 10000.
  6. Power supply - 2 battery AAA.
  7. Time of continuous operation from battery of Alkaline type, not less than 300 h.
  8. Dimensions, mm: 120x60x250.
  9. Gauge weight: 150g.

Particulars

  • Speed and high locality of measurements;
  • up to nine fixed material groups in the memory of the probe in the supply (at customer's request);
  • large number of probes with removable support rings for control the products varied destination - from large to small of profile composite;
  • a small influence of thickness and weight of products on the measurement results;
  • LED matrix with user-friendly interface and enhanced keyboard (based on the principle of one button - one function) for easy work with the device;
  • output on indicator mean and a single (final) results of measurements at the same time;
  • an opportunity to see the statistics in the process of measurement;
  • the possibility of tolerance control;
  • indicator back lights;
  • possibility of storing calibrations   in the memory of the probes and measurement results in memory of gauge with consequent transfer of then into IBM PC via channel USB2.0 for statistic processing, storage and documentation using the program «NDT1 - DATA - hardness».

 


Material

Scale

Scale dimension

Measurement range

Steel and cast steel

Ultimate strength sв

MPa

370…1740

Brinell

HB

75-650

Rockwell

HRC

20-70

Vickers

HV

75-1000

Rockwell "A"

HRA

70-93

Rockwell "B"

HRB

25-100

Super-Rockwell

HRN15

70-94

Super-Rockwell

HRN30

40-86

Super-Rockwell

HRN45

20-78

Super-Rockwell

HRT30

15-82

Shor

HSD

20-100

Leeb

HL

0-1000

Cast aluminum alloys

Brinell

HB

20-165

Rockwell "B"

HRB

24-85

Cold work tool steel

Vickers

HV

80-900

Rockwell

HRC

20-70

Stainless steel

Brinell

HB

85-650

Vickers

HV

85-800

Rockwell

HRC

20-60

Rockwell "B"

HRB

45-100

Gray cast iron

Brinell

HB

100-350

Vickers

HV

100-360

Rockwell

HRC

20-40

Nodular cast iron

Brinell

HB

150-400

Vickers

HV

150-420

Rockwell

HRC

20-42

Brass (copper-zinc alloys)

Brinell

HB

40-170

Rockwell "B"

HRB

14-95

Bronze (copper-aluminum/copper-tinalloys)

Brinell

HB

60-300

Wrought copper alloys

Brinell

HB

45-315

 

Dynamic probes for gauge KT

D/DC

  • dimensions, mm: Ø26x143/85;
  • roughness, mm: 3,2;
  • min metal thickness, mm: 10.

D+15

  • dimensions, mm: Ø26x158;
  • roughness, mm: 3,2;
  • min metal thickness, mm: 10.

C

  • dimensions, mm: Ø26x143;
  • roughness, mm: 3,2;
  • min metal thickness, mm: 10.

G

  • dimensions, mm: Ø33x268;
  • roughness, mm: 12,5;
  • min metal thickness, mm: 100.

 

Ultrasonic (UCI) probes for gauge KT

U-10N

Pressure Force 10 N.

  • min metal thickness, mm, 1;
  • min radius for test, mm, 5;
  • max surface roughness Ra, µm, 0,8 ;
  • dimensions Ø 26x130.

U-50N

Pressure Force 50 N.

  • min metal thickness, mm, 1;
  • min radius for test, mm, 5;
  • max surface roughness Ra, µm, 1,6 ;
  • dimensions Ø 26x130.

U-100N

Pressure Force 100 N.

  • min metal thickness, mm, 1,2;
  • min radius for test, mm, 5;
  • max surface roughness Ra, µm, 3,2 ;
  • dimensions Ø 26x130.
 

Package Contents

The gauge with the probes (and the number changed to your choice).

Delivery set

Electronic module, one probe, charger, batteries AAA -2 pc., passport,  calibration guide, cable of communication with PC and diskette with program of data transferring and processing of measurement data «NDT1-DATA-determination», packaging . If necessary, gauge can be completed with additional probes, hardness measures, and graduation of gauge for non-standard materials and articles can be made.

Warranty - 2 years.